Liquid Handling
114
Most Comfortable Pipetting The Acura ® manual fits everybody’s hand with or without glove. Ergonomic shape, light weight and soft plunger travel care for better comfort. Ejector Button Ergonomically positioned, the large surface, soft padded ejector provides for low-pressure activation.
Adjustable Tip Ejection Ejector shaft height adjustment system called Justip ™ allows a wide selection of tips to tightly fit the nozzle.
Minimum Maintenance / Sterilization Easy disassembling for thorough cleaning. Sterilization achieved by auto-claving instrument fully assembled (121°C / 250°F). Instant Calibration Swift-set user calibration system - with integrated key and locking mechanism - makes tedious procedures a thing of the past. Easy and precise, it is in full compliance with international standards and recommendations. Safety Seal Though easily removed when needed, the switch protection seal is autoclaving resistant.
Calibration Seal Stickers
Cat. No.
Description
Qty / Case
W810620 W810622
Replacement seals Replacement seals
25
100
Performance and Ordering Information - Acura ® manual 825 and 826 XS Micropipettes
Inaccuracy (E%)
Imprecision (CV%)
Cat. No.
Volume
Division
Min. Vol.
Mid. Vol.
Max. Vol.
Min. Vol.
Mid. Vol.
Max. Vol.
Weight (g)
Tip Style
Micropipettes Acura ® manual 825 W810300 0.1 - 2µL
0.002µL 0.01µL 0.01µL 0.02µL
< ± 6.0 % 1) < ± 2.5 % 2) < ± 2.5 % < ± 2.5 % < ± 1.5 % < ± 1.5 % < ± 1.5 % < ± 1.5 % < ± 6.0 % 1) < ± 2.5 % 2) < ± 2.5 % < ± 2.5 % < ± 1.5 % < ± 1.5 % < ± 1.5 % < ± 1.5 %
< ± 4.0 % < ± 1.8 % < ± 1.8 % < ± 1.8 % < ± 1.3 % < ± 1.2 % < ± 1.1 % < ± 1.0 % < ± 4.0 % < ± 1.8 % < ± 1.8 % < ± 1.8 % < ± 1.3 % < ± 1.2 % < ± 1.1 % < ± 1.0 %
< ± 2.0 % < ± 1.0 % < ± 1.0 % < ± 1.0 % < ± 1.0 % < ± 0.8 % < ± 0.6 % < ± 0.5 % < ± 2.0 % < ± 1.0 % < ± 1.0 % < ± 1.0 % < ± 1.0 % < ± 0.8 % < ± 0.6 % < ± 0.5 %
< 5.0 % 1) < 1.8 % 2)
< 3.3 % < 1.2 % < 1.6 % < 1.1 % < 0.7 % < 0.6 % < 0.4 % < 0.4 % < 3.3 % < 1.2 % < 1.6 % < 1.1 % < 0.7 % < 0.6 % < 0.4 % < 0.4 %
< 1.5 % < 0.5 % < 0.7 % < 0.5 % < 0.4 % < 0.2 % < 0.2 % < 0.2 % < 1.5 % < 0.5 % < 0.7 % < 0.5 % < 0.4 % < 0.2 % < 0.2 % < 0.2 %
88 88 88 88 90 92 90 95 84 84 85 85 86 87 86 88
10µL 10µL
W810302 W810304 W810306 W810308 W810310 W810312 W810314
0.5 - 10µL 1 - 10YµL*
< 2.5 % < 1.7 % < 1.0 % < 1.0 % < 0.6 % < 0.5 %
200µL 200µL 200µL 200µL 200µL 1000µL
2 - 20µL 5 - 50µL
0.1µL 0.1µL 0.2µL
10 - 100µL 20 - 200µL
100 - 1000µL 1µL
Micropipettes Acura ® manual 826 XS W810300-XS 0.1 - 2µL
0.002µL 0.01µL 0.01µL 0.02µL
< 5.0 % 1) < 1.8 % 2)
10µL 10µL
W810302-XS W810304-XS W810306-XS W810308-XS W810310-XS W810312-XS W810314-XS
0.5 - 10µL 1 - 10YµL*
< 2.5 % < 1.7 % < 1.0 % < 1.0 % < 0.6 % < 0.5 %
200µL 200µL 200µL 200µL 200µL 1000µL
2 - 20µL 5 - 50µL
0.1µL 0.1µL 0.2µL
10 - 100µL 20 - 200µL
100 - 1000µL 1µL
Performance values obtained with double-distilled water at constant temperature (± 0.5°C) comprised between 20°C and 25°C in accordance with ISO 8655. 1) measured at 0.5µL 2) measured at 1µL * This pipette must be used with a 200µL pipette tip