[DWK Life Sciences] Product Overview

Liquid Handling

114

Most Comfortable Pipetting  The Acura ® manual fits everybody’s hand with or without glove. Ergonomic shape, light weight and soft plunger travel care for better comfort. Ejector Button  Ergonomically positioned, the large surface, soft padded ejector provides for low-pressure activation.

Adjustable Tip Ejection  Ejector shaft height adjustment system called Justip ™ allows a wide selection of tips to tightly fit the nozzle.

Minimum Maintenance / Sterilization  Easy disassembling for thorough cleaning. Sterilization achieved by auto-claving instrument fully assembled (121°C / 250°F). Instant Calibration  Swift-set user calibration system - with integrated key and locking mechanism - makes tedious procedures a thing of the past. Easy and precise, it is in full compliance with international standards and recommendations. Safety Seal  Though easily removed when needed, the switch protection seal is autoclaving resistant.

Calibration Seal Stickers

Cat. No.

Description

Qty / Case

W810620 W810622

Replacement seals Replacement seals

25

100

Performance and Ordering Information - Acura ® manual 825 and 826 XS Micropipettes

Inaccuracy (E%)

Imprecision (CV%)

Cat. No.

Volume

Division

Min. Vol.

Mid. Vol.

Max. Vol.

Min. Vol.

Mid. Vol.

Max. Vol.

Weight (g)

Tip Style

Micropipettes Acura ® manual 825 W810300 0.1 - 2µL

0.002µL 0.01µL 0.01µL 0.02µL

< ± 6.0 % 1) < ± 2.5 % 2) < ± 2.5 % < ± 2.5 % < ± 1.5 % < ± 1.5 % < ± 1.5 % < ± 1.5 % < ± 6.0 % 1) < ± 2.5 % 2) < ± 2.5 % < ± 2.5 % < ± 1.5 % < ± 1.5 % < ± 1.5 % < ± 1.5 %

< ± 4.0 % < ± 1.8 % < ± 1.8 % < ± 1.8 % < ± 1.3 % < ± 1.2 % < ± 1.1 % < ± 1.0 % < ± 4.0 % < ± 1.8 % < ± 1.8 % < ± 1.8 % < ± 1.3 % < ± 1.2 % < ± 1.1 % < ± 1.0 %

< ± 2.0 % < ± 1.0 % < ± 1.0 % < ± 1.0 % < ± 1.0 % < ± 0.8 % < ± 0.6 % < ± 0.5 % < ± 2.0 % < ± 1.0 % < ± 1.0 % < ± 1.0 % < ± 1.0 % < ± 0.8 % < ± 0.6 % < ± 0.5 %

< 5.0 % 1) < 1.8 % 2)

< 3.3 % < 1.2 % < 1.6 % < 1.1 % < 0.7 % < 0.6 % < 0.4 % < 0.4 % < 3.3 % < 1.2 % < 1.6 % < 1.1 % < 0.7 % < 0.6 % < 0.4 % < 0.4 %

< 1.5 % < 0.5 % < 0.7 % < 0.5 % < 0.4 % < 0.2 % < 0.2 % < 0.2 % < 1.5 % < 0.5 % < 0.7 % < 0.5 % < 0.4 % < 0.2 % < 0.2 % < 0.2 %

88 88 88 88 90 92 90 95 84 84 85 85 86 87 86 88

10µL 10µL

W810302 W810304 W810306 W810308 W810310 W810312 W810314

0.5 - 10µL 1 - 10YµL*

< 2.5 % < 1.7 % < 1.0 % < 1.0 % < 0.6 % < 0.5 %

200µL 200µL 200µL 200µL 200µL 1000µL

2 - 20µL 5 - 50µL

0.1µL 0.1µL 0.2µL

10 - 100µL 20 - 200µL

100 - 1000µL 1µL

Micropipettes Acura ® manual 826 XS W810300-XS 0.1 - 2µL

0.002µL 0.01µL 0.01µL 0.02µL

< 5.0 % 1) < 1.8 % 2)

10µL 10µL

W810302-XS W810304-XS W810306-XS W810308-XS W810310-XS W810312-XS W810314-XS

0.5 - 10µL 1 - 10YµL*

< 2.5 % < 1.7 % < 1.0 % < 1.0 % < 0.6 % < 0.5 %

200µL 200µL 200µL 200µL 200µL 1000µL

2 - 20µL 5 - 50µL

0.1µL 0.1µL 0.2µL

10 - 100µL 20 - 200µL

100 - 1000µL 1µL

Performance values obtained with double-distilled water at constant temperature (± 0.5°C) comprised between 20°C and 25°C in accordance with ISO 8655. 1) measured at 0.5µL 2) measured at 1µL * This pipette must be used with a 200µL pipette tip