[DWK Life Sciences] Product Overview

Liquid Handling

120

Calibra ® digital 822 Micro / 832 Macro Pipettes The combination of instant volume adjustment, mechanical precision and ease of use makes the Calibra ® digital a most reliable and robust

micropipette line in the market. Product Advantages ■ Winding-free, instant, volume adjustment ■ Long term performance and calibration stability ■ Robust, long lasting construction ■ Shock, UV light and autoclave resistant ■ Reliable user calibration ■ Minimal maintenance requirements ■ CE certified IVD 98/79 EEC ■ Interchangeable nozzle filter in all macro models ■ Three-year warranty

822 micro 0.2 - 2µL 1 - 10µL 2 - 20µL 10 - 100µL 20 - 200µL 100 - 1000µL

832 macro 0.2 - 2mL 1 - 10mL

Thin shaft shape helps reach contents from microtubes and/ or narrow containers. The Calibra ® provides convenient access into those hard to reach places.

Instant Volume Setting Twin cam system fitted with pre-calibrated steps providing for true digital volume entry and numerical display. The dual incrementation allows fast volume setting without tedious winding.

Key Volume Adjustment Example: from 10 to 50µL in half a revolution of the setting wheel placed in its normal position.

Fine Tuning Adjustment Example: from 50 to 55.5µL in half a revolution of the setting wheel in its “pulled” position.

Performance and Ordering Information

Inaccuracy (E%)

Imprecision (CV%)

Cat. No.

Volume

Division

Min. Vol.

Mid. Vol.

Max. Vol.

Min. Vol.

Mid. Vol.

Max. Vol.

Tip Style

Calibra ® digital 822 851160

0.2 - 2µL 1 - 10µL 2 - 20µL

0.1µL

< ± 6.0 % 1) < ± 2.5 % < ± 2.5 % < ± 1.5 % < ± 1.5 % < ± 1.5 %

< ± 4.0 % < ± 1.5 % < ± 1.5 % < ± 0.9 % < ± 0.9 % < ± 0.6 %

< ± 2.0 % < ± 1.0 % < ± 1.0 % < ± 0.8 % < ± 0.6 % < ± 0.5 %

< 5.0 % 1)

< 3.3 % < 1.3 % < 1.1 % < 0.6 % < 0.4 % < 0.4 %

< 1.5 % < 0.5 % < 0.5 % < 0.2 % < 0.2 % < 0.2 %

10µL 10µL

851162 851163 851164 851166 851168

0.05µL

< 2.0 % < 1.7 % < 1.0 % < 0.6 % < 0.5 %

0.1µL 0.5µL 1.0µL 5.0µL

200µL 200µL 200µL 1000µL

10 - 100µL 20 - 200µL

100 - 1000µL

Calibra ® digital 832 851340

0.2 - 2mL 1 - 10mL

0.01mL 0.1mL

< ± 1.5 % < ± 1.5 %

< ± 1.0 % < ± 1.0 %

< ± 0.5 % < ± 0.5 %

< 0.5 % < 0.3 %

< 0.3 % < 0.3 %

< 0.2 % < 0.15 %

2mL

851345

10mL

Performance values obtained with double-distilled water at constant temperature (± 0.5°C) comprised between 20 and 25°C in accordance with ISO 8655. 1) measured at 0.5µL

Powered by