KINETIC CHROMOGENIC METHOD RECOMBINANT REAGENT
PyroSmart NextGen ® Recombinant Cascade Reagent (rCR)
Sample to Lysate Ratio PyroSmart NextGen ® is used at with an economical vol- ume of 50 µL of reagent per well yielding 50 tests/vial:
• Microplate reader: 1:1 ratio using 50 µL of test sample : 50 µL of reagent
Test Performance The PyroSmart NextGen ® reaction mixture is incubated at 37±1°C and read in a microplate reader equipped with a 405–410 nm filter. The time of incubation is dependent on the lowest standard concentration in the standard curve, with 0.005 EU/mL achievable in 2500 seconds in a microplate reader. Software is used to construct the standard curve and calculate the endotoxin concentrations. Product Stability PyroSmart NextGen ® is a lyophilized product with an excellent shelf life of 3 years from the date of manufacture. Product Reconstitution PyroSmart NextGen ® is provided as co-lyophilized with the chromogenic substrate and as such it is ready-to-use following a simple reconstitution (using 2.8 mL of the supplied reconstitution buffer). Product Packaging PyroSmart NextGen ® reagent is provided as a pack of 2 vials of reagent and 2 vials of reconstitution buffer. This is sufficient for a total of 110 wells (55 wells per vial).
PyroSmart NextGen ® recombinant Cascade Reagent (rCR) marks the introduction of a new sustainable recombinant LAL reagent technology for Bacterial Endotoxin Testing (BET). Utilizing the same LAL cascade as traditional LAL reagents, while eliminating the potential for (1 g 3)- b - D -Glucans cross reactivity, PyroSmart NextGen ® delivers all of the quality and consistency of results you have come to expect from ACC LAL reagents. PyroSmart NextGen ® can be used for a wide variety of endotoxin tests, ranging from standard water testing to samples requiring high sensitivity, such as intrathecal products and those requiring high dilutions to overcome interference. Product Sensitivity The sensitivity for recombinant chromogenic assays is determined by the lowest standard concentration on the standard curve used for the assay. The maximum sensitivity of PyroSmart NextGen ® is 0.005 EU/mL when run in an incubating microplate reader (or 0.001 EU/mL when run in Pyros Kinetix ® Flex tube reader).
Multi-Test 2.8 mL/vial (approx. 50 tests/vial) #PNG050-2 Each kit contains
2 x PyroSmart NextGen ® Reagent 2 x Reconstitution Buffer
17
Powered by FlippingBook