CYTOTOXIC DRUG SAFETY CABINETS
Laminar air flow pattern through a Cytotoxic cabinet.
SPECIFICATIONS
OPERATION Filters HEPA
Mk 5 Series 1200mm CDSC
Mk 5 Series 1800mm CDSC
99.997% to 0.3 micron 800 Lux at work surface
99.997% to .3 micron 800 Lux at work surface
Lighting
Germicidal UV lamp
400mW/m 2 240v/50 Hz
400mW/m 2 240v/50 Hz
Electrical supply
Power Consumption (Kw)
1
1.5
Fans
3 x 240V Single Phase Direct Drive
3 x 240V Single Phase Direct Drive
Exhaust air volume (m 3 /Hr)
1044
1548
Noise Level
65dBA at 1m
65dBA at 1m
Left, Side, Right or Top
Top only
Discharge Options DIMENSIONS Work Aperture
210mm
190mm
Work area W x D x H (mm)
1180 x 560 x 620 1345 x 750 x 2216
1760 x 560 x 620 1990 x 750 x 2380
Outer dimensions W x D x H (mm)
Weight ((kg)
360
630
CONSTRUCTION Front screen Internal work area & tray Outer construction Manufacturing Compliance to
6mm Laminated glass 304 Stainless Steel Mild steel with powder coating
AS 2567 - 2000
Certification to
AS1807.1 / AS1807.5 / AS1807.6 / AS1807.15 / AS1807.20 / AS1807.22 / AS1807.23
Powered by FlippingBook