[Copley Scientific] Product Overview

Dissolution

DISi Series: Technical Specifications Pharmacopoeial Compliance

Ph. Eur. Chapters: 2.9.3, 2.9.4 USP Chapters: <711> <724>

User Interface

Resistive touchscreen

Speed Range

20 - 220 rpm +/- 2%

Heater Accuracy

± 0.1°C

Heater Temperature Range

Ambient - 50°C

Vibration Elimination

Low vibration heaters

Test Run Time

Up to 99 hours, 59 minutes, 59 seconds

Alarm(s)

1 . End of testing (audible)

2 . Low bath water level warning on-screen indicator

Data Output

RS 232 USB A (for connection with a USB printer) USB B (for connection with a PC)

Temperature Calibration

Single-point electronic temperature calibration . Calibration of the DISi Series temperature probe is simple, through the use of an electronic calibration key and passcode-protected calibration menu

designed to guide users through the process without fuss. The latest temperature probe calibration information is stored and available to print/export when convenient to the user.

Electronic Temperature Calibration Key

DISi Series

Cat. No.

Description

1336 1338 1209 1309 1307

Dissolution Tester DIS 600i (incl. 6 Drive Shafts) Dissolution Tester DIS 800i (incl. 8 Drive Shafts)

Electronic Temperature Calibration Key

IQ/OQ/PQ Documentation Pack Printer (including USB cable)

31